首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Contrast-based ultrasonic blood flow measurements based on inflow/outflow time intensities
【24h】

Contrast-based ultrasonic blood flow measurements based on inflow/outflow time intensities

机译:基于对比的超声波血流测量基于流入/流出时间强度

获取原文

摘要

Ultrasonic contrast specific blood flow measurement techniques offer new opportunities to assess blood flow information based on evaluation of time-intensity curves (TICs). Such curves are measured to estimate concentration of the microbubbles in the blood pool. Based on the indicator-dilution theory, hemodynamic parameters such as the volume flow rate can be directly related to the time constant estimated from the TICs. In this paper, the applicability of the indicator-dilution theory is studied with an in vitro experiment setup. Moreover, the linear system assumption of the blood mixing mechanism is also tested. Several flow phantoms are constructed and a self-made, albumin based contrast agent is used. The TIC is measured by using B-mode images obtained from a commercial ultrasound system. It is found that with a bolus injection and a single mixing chamber, the estimated time constants agree with the theory despite that the effective mixing volume may be smaller than the actual mixing chamber volume in some conditions. More importantly, discrepancy also exists with a prolonged injection and/or two mixing chambers with cascade connection. In other words, the linear system assumption is still questionable even under the controlled in vitro experimental conditions. Potential sources of the discrepancy require further investigation in order to develop contrast specific quantitative blood flow measurement techniques.
机译:超声波对比度特定血流测量技术提供了基于时间强度曲线评估(TICS)评估血流信息的新机会。测量这种曲线以估计血液池中微泡的浓度。基于指示剂稀释理论,血流动力学参数如体积流量可以与来自TIC估计的时间常数直接相关。本文用体外实验设置研究了指标稀释理论的适用性。此外,还测试了血液混合机构的线性系统假设。构建了几种流动模型,使用自成的白蛋白基于白蛋白的造影剂。通过使用从商业超声系统获得的B模式图像来测量TIC。发现,通过推注喷射和单个混合室,估计的时间常数仍然认为,尽管有效混合体积可以小于一些条件下的实际混合室体积。更重要的是,差异也存在于具有级联连接的延长的注射和/或两个混合室。换句话说,即使在受控的体外实验条件下,线性系统假设仍然是可疑的。差异的潜在来源需要进一步调查以发展对比特异性定量血流测量技术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号