A study was conducted to examine the use of the MiniMed Continuous Glucose Monitoring System (COMS) to measure blood sugar level (BSL) in real-time. Eight patients participated in the study after informed consent was obtained. When more than a 3 mmol/L deviation occurred between CGMS reading and BSL as determined by conventional blood glucose meter using blood sample drawn from an arterial line, CGMS was re-calibrated, and "regression calibration" was applied to better estimate the real-time BSL. The mean difference between the CGMS and Meter BSL readings was found to be -0.2 ± 5.3 mmol/L (Mean ± 1.96 × Standard Deviation). Clarke's Error grid analysis showed that 64.6% of the CGMS BSL readings were clinically accurate (i.e. <20% deviation from Meter BSL) while 30.9% were clinically acceptable. A calibration technique suitable for use in estimating BSL in real-time is needed to make CGMS useful in clinical situations.
机译:进行了研究以检查最小连续葡萄糖监测系统(COMS)的使用实时测量血糖水平(BSL)。获得了八名患者在获得知情同意后参加了研究。当CGMS读数和BSL之间发生超过3mmol / L偏差时,使用从动脉线汲取的血液样本测定的血葡萄糖仪,CGM被重新校准,并施加“回归校准”以更好地估计真实 - 时间bsl。发现CGM和仪表BSL读数的平均差异为-0.2±5.3mmol / L(平均±1.96×标准偏差)。 Clarke的误差网格分析表明,64.6%的CGMS BSL读数是临床准确的(即,从仪表BSL偏差),而30.9%在临床上可接受。需要一种适用于估计BSL实时BSL的校准技术,以使CGM在临床情况下可用。