Neutron spectra and the ambient dose equivalent were calculated inside the bunker of a 15 MV Varian linac model CLINAC iX. Calculations were carried out using Monte Carlo methods. Neutron spectra in the vicinity of isocentre show the presence of evaporation and knock-on neutrons produced by the source term, while epithermal and thermal neutron remain constant regardless the distance respect to isocentre, due to room return. Along the maze neutron spectra becomes softer as the detector moves along the maze. The ambient dose equivalent is decreased but do not follow the 1 /r~2 rule due to changes in the neutron spectra.