Comparisons of the Codes of Practice IAEA TRS 277 and TRS 398: High Energy Photons and Electron Beams

机译:实践代码的比较IAEA TRS 277和TRS 398:高能量光子和电子束

获取原文

摘要

This work presents the calibration for: 6, 8, 10, 12, 14, 15 and 18 MeV electron beams, also to 6 and 15 MV photons beams. Beams that are generated by an accelerator Elekta Linac SL 18. The calibration is performed in terms of absorbed dose to water D_w. It is determined by two different protocols: the code of practice of the International Atomic Energy Agency (IAEA) TRS 277 for ionization chambers calibrated on air kerma N_K, and the code of practice TRS 398 for ionization chambers calibrated on absorbed dose to water N_(D,W). Two independent dosimeters were used with two ionization chambers each one, respectively. The first one, a dosimeter PTW model UNIDOS with Markus type chamber (plane parallel) for electrons, and Farmer type chamber (cylindrical) for photons, both chambers calibrated in N_K. The second dosimeter Scanditronix model DOSE 1 with plane parallel chamber (electrons) and cylindrical chamber (photons), both chambers calibrated in terms of N_(D,W). In the case of photon beams, the TPR was measured for 6 and 15 MV, also the profiles were determined in order to verify the flatness and symmetry of the beam: ±3%. The quality for electrons beams were estimated by means of the D_(max), R_(80), R_(50) and R_p. The results obtained for the absorbed dose quotients D_(W,Q)(d_(max))_(277)~(398) are: Electrons, 0.976 ≤ D_(W,Q)(d_(max))_(277)~(398) ≤ 1.001; Photons: 1.001 ≤ D_(W,Q)(d_(max))_(277)~(398) ≤ 1.003.
机译:该工作呈现:6,8,10,12,14,15和18MeV电子束,也呈现为6和15mV光子束。由加速器Elekta Linac S118产生的光束。校准在吸收剂量至水D_W方面进行。它是由两种不同的协议决定:国际原子能机构(IAEA)的工作守则TRS 277上的空气比释动能N_K校准电离室,和实践的代码TRS 398在吸收剂量校准,以水N_电离室( d,w)。两个独立剂量计分别与两个电离室一起使用。第一个,一个剂量计PTW模型UniDOS,用于电子的Markus型室(平面),以及用于光子的农民式腔室(圆柱形),两个腔室都校准在N_K中。具有平面平行室(电子)和圆柱形腔室(光子)的第二剂表钪型模型1,两个腔室以N_(D,W)校准。在光子束的情况下,测量TPR 6和15mV,确定曲线以验证光束的平坦度和对称性:±3%。通过D_(MAX),R_(80),R_(50)和R_P估计电子束的质量。对于所获得的结果的吸收剂量商数D_(W,Q)(D_(最大值))_(277)〜(398)为:电子,0.976≤D_(W,Q)(D_(最大值))_(277) 〜(398)≤1.001;光子:1.001≤d_(w,q)(d_(max))_(277)〜(398)≤1.003。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号