首页> 外文会议>Mexican Symposium on Medical Physics >Pilot Quality Control Program for Audit RT External Beams at Mexican Hospitals
【24h】

Pilot Quality Control Program for Audit RT External Beams at Mexican Hospitals

机译:墨西哥医院审计外置梁的试点质量控制计划

获取原文

摘要

A pilot quality control program for audit 18 radiotherapy RT external beams at 13Mexican hospitals is described --for eleven ~(60)Co beams and seven photon of 6, 10 15MV from accelerators. This program contains five parts: a) Preparation of the TLD-100 powder:washing, drying and annealing (one hour 400 °C plus 24 hrs 80 °C). b) Sending two IAEA typecapsules to the hospitals for irradiation at the hospital to a nominal D_w= 2 Gy. c) Preparation at theSSDL of ten calibration curves CC in the range of 0.5 Gy to 6 Gy in terms of absorbed dose towater Dw for ~(60)Co with traceability to primary laboratory NRC (Canada), according to a windowirradiation: 26/10/2007-7/12/2007. d) Reading all capsules that match their hospital timeirradiation and the SSDL window irradiation. f) Evaluation of the Dw imparted by the hospitals.
机译:13mexican医院的审计18个放射疗法RT外梁的试验质量控制程序 - 对于11,来自加速器的七个光子,七个光子为6,1015mV。该程序包含五个部分:a)制备TLD-100粉末:洗涤,干燥和退火(1小时400°C加24小时80°C)。 b)向医院发送两款IAEA的TYPEAPLE,以在医院照射到名义D_W = 2 GY。 c)根据WindowAimiratiation:26/10,在1(60)CO的〜(60)CO的吸收剂量拖曳DW方面,在10次校准曲线CC的制备在0.5 GY至6 GY的范围内。 / 2007-7 / 12/2007。 d)阅读与其医院时间内的所有胶囊和SSDL窗射线照射。 f)评估医院赋予的DW。