CZT detector for dual-energy x-ray absorptiometry (DEXA)

机译:用于双能X射线吸收度(DEXA)的CZT检测器

获取原文

摘要

A cadmium-zinc-telluride (CZT) detector has been developed for a bone densitometer that uses dual-energy x-ray absorptiometry (DEXA) to determine bone mineral density in vivo. A linear array of 16 discrete CZT detectors is used with a narrow fan-shaped x-ray beam to scan the patient. Each detector is 3 mm thick and 7 mm * 3 mm in area and has simple planar contacts. The x-ray beam has two broad energy lobes with effective energies of ~38 keV and ~65 keV. The energy sensitivity of the CZT detectors allows discrimination between low and high energy x-rays. Using DEXA, the relative difference in these two count rates permits a quantitative measurement of the areal densities of bone mineral and soft tissue. The detectors demonstrate good performance characteristics and stable operation in a clinical environment. This paper discusses the suitability of CZT for use in DEXA applications and describes its successful implementation and performance in this bone densitometer.
机译:已经开发了一种镉 - 锌 - 碲化物(CZT)探测器用于骨密度计,用于使用双能X射线吸收术(DEXA)来确定体内骨密度的骨密度。 16个离散CZT探测器的线性阵列与窄扇形X射线束一起使用以扫描患者。每个探测器厚,面积为3毫米厚,7毫米* 3毫米,具有简单的平面触点。 X射线束有两个宽的能量裂隙,具有〜38 keV和〜65 kev的有效能量。 CZT检测器的能量灵敏度允许低能量X射线之间的歧视。使用德克萨,这两种计数率的相对差异允许定量测量骨矿物和软组织的面部密度。探测器在临床环境中表现出良好的性能特征和稳定的操作。本文讨论了CZT在DEXA应用中使用的适用性,并描述了在这种骨密度计中的成功实施和性能。