A novel method for analog measuring of pulse width and rise time of defibrillating shocks

机译:一种新型脉冲宽度测量方法和除颤冲击的上升时间

获取原文

摘要

A novel method and the instrumentation developed for analog measuring of pulse width (T{sub}10 or T{sub}50) and rise time (T{sub}(10-90)) of cardiac defibrillators discharges are presented. The input pulse is applied to an analog delay network and to a peak detector which determines reference voltages. Those phase-shifted signals go to a comparator that generates the control pulse. A counter and a decoder excite the LCD display that shows the measured parameters with accuracy better than 0.33%.
机译:提出了一种新的方法和用于模拟测量的脉冲宽度(T {Sub} 10或T {Sub} 50)和上升时间(T {Sub}(10-90)的模拟测量的仪器。输入脉冲被施加到模拟延迟网络和峰值检测器,该峰值检测器确定参考电压。这些相移信号转到生成控制脉冲的比较器。计数器和解码器激励LCD显示屏,其表示测量参数,精度优于0.33%。