首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Correction of time-intensity curves for tissue perfusion evaluation using ultrasound contrast agent
【24h】

Correction of time-intensity curves for tissue perfusion evaluation using ultrasound contrast agent

机译:超声造影剂校正组织灌注评价的时间强度曲线

获取原文

摘要

Instability of scattering of ultrasound contrast agents (UCA) during their interaction with flowing blood will modify their time-intensity curves, which imposes a limitation on the ability of time-intensity curves for UCA aided tissue perfusion evaluation. A direct correction method, using the varying scattering of UCA as a function of time, is proposed and verified with an in vitro flow model and in vivo studies on six anesthetized open-chest dogs. The result of the experiment demonstrates that the homemade UCA is efficient, and the parameters derived from the corrected time-intensity, especially areas under the curve, are more closely correlated with myocardial ischemic degree. This method appears to be a useful addition to existing clinical techniques for quantifying tissue perfusion with UCA.
机译:在其与流动血液相互作用期间超声造影剂(UCA)散射的不稳定性将改变它们的时间强度曲线,这对UCA辅助组织灌注评估的时间强度曲线的能力提出了限制。用六个麻醉露胸部的体外流动模型提出并验证了使用UCA的变化散射的直接校正方法。实验结果表明,自制UCA是有效的,并且源自校正的时间强度,尤其是曲线下的区域的参数与心肌缺血程度更密切相关。该方法似乎是现有临床技术的有用的补充,用于量化UCA的组织灌注。