An implantable telemetry system, using a capacitor having high capacitance instead of the rechargeable batteries, has been designed for chronic unanaesthetized animal studies. This system can simultaneously record the cardiac sympathetic nerve activity and ECG. The power of the system is inductively coupled and transmitted to a 2.5 farad capacitor. The system, which is constructed by ultralow power analog and digital standard integrated circuits, has a transmission range up to 5 m, depending on intervening tissue depth and FM receiver characteristics. With the capacitor charging time of 5 minutes, the system can be driven for 120 minutes.
机译:使用具有高电容代替可充电电池的电容器的可植入遥测系统已经专为慢性非展开的动物研究设计。该系统可以同时记录心脏交感神经活动和心电图。系统的功率是电感耦合并传输到2.5前电容器。由超级功率模拟和数字标准集成电路构造的系统具有高达5米的传输范围,这取决于中间组织深度和FM接收器特性。随着电容器充电时间为5分钟,系统可以驱动120分钟。