首页> 外文会议>Conference on EUV,X-Ray,and Gamma-Ray Instrumentation for Astronomy >Use of x-ray CCDs in the calibration of the flight mirrors for the JET-X telescope on Spectrum-X
【24h】

Use of x-ray CCDs in the calibration of the flight mirrors for the JET-X telescope on Spectrum-X

机译:在Spectrum-X上校准Jet-X望远镜的飞行镜中的X射线CCD

获取原文

摘要

The x-ray mirror calibration program for the JET-X telescope on spectrum-X has recently been carried out at the 130 m long Panter x-ray beam line of the Max Plank Institute fur Extraterrestriche Physik. The excellent spatial resolution achieved with these mirrors, 15 arcsec half energy width (HEW) at 1.5 keV and 19 arcseconds at 8 keV, has proved to be difficult to measure precisely using previously established calibration methods (involving either slit detectors or the ROSAT PSPC imaging proportional counter). New diagnostic techniques have, therefore, been developed using a CCD imaging camera which utilized newly available x- ray CCD technology. Details of the calibration technique and the performance of the camera are provided and results are compared with those obtained from the slit and PSPC detectors.
机译:最近在Max Plank Institute Fur Extristriche Peavyik的130米长的肠道X射线束线上进行了X射线镜校准程序的X射线镜校准程序。通过以前8keV的1.5keV和19个弧形秒为10个弧形半能量宽度(HEW)实现的优异空间分辨率已经证明难以使用先前建立的校准方法(涉及狭缝检测器或ROSAT PSPC成像)精确测量比例计数器)。因此,新的诊断技术已经使用了使用新可用的X射线CCD技术的CCD成像摄像头开发。提供了校准技术的细节和相机的性能,并将结果与​​从狭缝和PSPC检测器获得的结果进行比较。