首页> 外文会议>Annual scientific meeting of computers in cardiology >Single beat spectral variance (SBSV): a new method to investigate both constant and time-varying ventricular late potentials
【24h】

Single beat spectral variance (SBSV): a new method to investigate both constant and time-varying ventricular late potentials

机译:单节拍光谱差异(SBSV):一种研究常数和时变心室晚期潜力的新方法

获取原文

摘要

Signal averaging (SA) extinguishes variable late potentials (LP). This might disclose important information for risk stratification of patients prone to ventricular tachycardia (VT). SBSV uses 2-dimensional fast Fourier transform (2D-FFT) to evaluate beat-to-beat variabilities. 2D-FFT shows the frequency contents in the first dimension and the beat-to-beat periodicity in the second dimension. A LP-index function allows the detection of constant and variable LPs. 85 patients (P) after myocardial infarction were investigated: 35 P with VT (group 1) and 50 P without VT (group 2). With SBSV 29/35 P (83%) revealed LP in group 1, 14 of 29 with marked Wenckebach periodicity, and only 5/50 P (10%) in group 2. In comparison, the Simson method after signal averaging detected LP in 18/35 P (51%) in group 1 and 9/50 P (18%) in group 2. Thus, SBSV demonstrated, that variable appearance of LP might be an explanation for negative results after signal averaging.
机译:信号平均(SA)熄灭可变后期电位(LP)。这可能会披露患者易受心室性心动过速(VT)的患者风险分层的重要信息。 SBSV使用二维快速傅里叶变换(2D-FFT)来评估节拍变化的变换。图2D-FFT显示了第一维的频率内容和第二维中的节拍周期性。 LP-index功能允许检测常数和可变LPS。研究了85名患者(P)在心肌梗塞后进行了研究:35 p,VT(1)和50p,没有VT(第2组)。随着SBSV 29/35 P(83%)在第1组,第29组中的LP揭示了标记的Wenckebach周期,并且在第2组中仅5/50 p(10%)。相比之下,信号平均检测到LP后的Simson方法18/35p(51%)在第2组中的第1组和9/50 p(18%)中,SBSV证明,LP的可变外观可能是信号平均后的负面结果的解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号