Minute Roughness Measurement Using Phase Tracking for Arterial Wall Diagnosis Non-Invasively in vivo

机译:微小粗糙度测量使用相位跟踪进行动脉壁诊断非侵入性体内

获取原文

摘要

Early diagnosis of atherosclerosis is essential as many of the risk factors are life-style dependent. We suggest a new method to measure minute roughness of the size of micrometers of the arterial wall. The method was evaluated using three silicone phantoms sized 13μm, 23μm and 33μm, respectively. The mean of the measured heights of the phantoms were 8.1μm (SD 0.0), 23.3μm (SD 0.2) and 29.6μm (SD 0.1) in the forward direction, and 7.7μm (SD 0.0), 21.9μm (SD 0.2) and 27.3μm (SD 3.6) in the backward direction, respectively. The phantom study shows very promising result and encourages to further evaluation and in vivo investigations.
机译:早期诊断动脉粥样硬化是必不可少的,因为许多风险因素都是生活方式依赖。我们建议一种测量动脉墙的微米尺寸的微小粗糙度的新方法。使用三个硅氧烷幻影为13μm,23μm和33μm评估该方法。模薄的测量高度的平均值在向前方向为8.1μm(SD 0.0),23.3μm(SD 0.2)和29.6μm(SD 0.1),7.7μm(SD 0.0),21.9μm(SD 0.2)和分别在向后方向27.3μm(SD 3.6)。幻影研究表明了非常有希望的结果,并鼓励进一步评估和体内调查。