摘要

A 680kHz confocal airborne ultrasonic instrument was built for quantitative material characterization based on parametric imaging. This instrument allows characterization of fragile or moist sensitive samples. Few such instruments are commercially available. The instrument performs simultaneous attenuation and time-of-flight measurements in linear reflection and transmission modes. Sample surface topography is measured in reflection mode and local sample thickness is calculated using the inter-transducer distance. Results for a Pinus Sylvestris test sample are presented. The speed of sound, attenuation, and modulus of elasticity can be mapped locally (local density was measured with X-ray in transmission mode). No repeatable nonlinear sample characteristics were achieved for this sample.
机译:基于参数成像的定量材料表征,建立了680kHz共聚焦空中超声仪。该仪器允许表征易碎或湿润的敏感样品。很少有这些仪器可商购获得。该仪器在线性反射和传输模式下执行同时衰减和飞行时间测量。样品表面形貌在反射模式下测量,并且使用换体距离计算局部样品厚度。提出了Pinus Sylvestris测试样品的结果。可以在本地映射声音,衰减和弹性模量(在传输模式下用X射线测量局部密度)。对于该样品,没有实现可重复的非线性样品特性。