Hemodialysis is the most frequently used therapy for people with chronic kidney failure. However, there are several obstacles in hemodialysis therapy, first the formation of blood clots when the blood is pumped out of the body, then the speed of pumping blood is not stable causing the patient to experience in shock. This research focuses on design peristaltic pump and syringe pump using a motor stepper, by calibrating the amount step motor stepper to keep blood flow out stable at 200 ml/min and control the heparin injection so that blood clots do not form. The designed peristaltic pump can produce flow rate close to 200 ml/min and have levels with a precision of 98.4899% and an accuracy rate of 98.6155%. Syringe pump which is designed to inject heparin can work well. The resulting the precision level is 99.8797% and accuracy rate is 99.3861%.
机译:血液透析是慢性肾功能衰竭的人最常使用的疗法。然而,血液透析治疗中有几个障碍,首先形成血液凝血时血液被泵出体内,然后泵送血液的速度不稳定,导致患者休克体验。本研究专注于设计蠕动泵和注射器泵使用电机步进,通过校准量步进电机步进器,以保持血液流出稳定,并控制肝素注射,使血凝块不形成血栓。设计的蠕动泵可以产生接近200毫升/分钟的流速,精度为98.4899%,精度率为98.6155%。设计用于注射肝素的注射器泵可以很好地工作。得到的精确度为99.8797%,精度率为99.3861%。