首页> 外文会议>International Conference on Hard X-Ray, Gamma-Ray, and Neutron Detector Physics >Developing a dual energy x-ray absorptiometry (DEXA) system using SrI2:Eu~(2+) coupled to silicon photomultiplier (SiPM)
【24h】

Developing a dual energy x-ray absorptiometry (DEXA) system using SrI2:Eu~(2+) coupled to silicon photomultiplier (SiPM)

机译:使用SRI2的双能X射线吸收计(DEXA)系统:EU〜(2+)耦合到硅光电倍增管(SIPM)

获取原文

摘要

This paper presents the evaluation of an alternative solution for detectors of dual energy x-ray absorptiometry systems. Bone mineral density (BMD) evaluation tools are crucial to the proper diagnosis of osteoporosis. The commercially available DEXA systems utilize CZT as the detector of the small dose of ionizing radiation passing through the area of interest. Our paper presents a novel Strontium Iodide doped with Europium (SrI2:Eu~(2+)) scintillator crystal coupled to a Silicon Photomultiplier (SiPM) array as a less expensive alternative to CZT detector. Dual energy (60 keV and 122 keV) exposure was used for BMD measurements of a bone phantoms containing CaHPO_4 to mimic bone and resin to mimic soft tissue. Calibration curves was determined for both BMD and for thickness measurements, validated by "unknown" phantoms. Both calibration curves lead to decreased errors compared to commercially available systems.
机译:本文介绍了双能X射线吸收系统检测器替代解决方案的评价。骨密度(BMD)评估工具对骨质疏松症的适当诊断至关重要。市售的DEXA系统利用CZT作为通过感兴趣区域的小剂量电离辐射的检测器。我们的论文提出了一种掺杂的新型锶碘化物(SRI2:Eu2:Eu〜(2+))闪烁体晶体,耦合到硅光电倍增器(SiPM)阵列,作为CZT检测器的较便宜的替代方案。双能量(60keV和122keV)曝光用于含CaHPO_4的骨模子的BMD测量以模拟骨和树脂以模拟软组织。测定BMD和用于厚度测量的校准曲线,由“未知”幽灵验证。与市售的系统相比,这两个校准曲线都导致误差降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号