首页> 外文会议>International Symposium on Medical Measurements and Applications >Temperature Monitoring during Radio Frequency Thermal Ablation Treatment on Ex Vivo perfused organ by Fiber Bragg Grating Sensors
【24h】

Temperature Monitoring during Radio Frequency Thermal Ablation Treatment on Ex Vivo perfused organ by Fiber Bragg Grating Sensors

机译:光纤光栅传感器对离体灌注器官进行射频热消融治疗期间的温度监测

获取原文

摘要

In this work, we report on the integration of fiber Bragg gratings sensors in a commercial radio frequency probe for real-time temperature monitoring during thermal ablation of biological tissues. The experiments were conducted on ex-vivo pig liver and bovine kidney tissues and the results confirm that the proposed setup is able to conduct a multi-point real-time temperature monitoring during the RF discharges with a temperature resolution of 0.1 °C and a minimum spatial resolution of 1 mm. During these experiments the blood perfusion was implemented by injecting water at a temperature of 37 °C inside the organ under examination. This solution aimed at simulating the Radio Frequency Thermal Ablation (RFTA) treatments made in clinical practice.
机译:在这项工作中,我们报告了光纤布拉格光栅传感器在商业射频探头中的集成,用于在生物组织热消融期间进行实时温度监测。实验是在离体猪肝和牛肾组织上进行的,结果证实了所提出的设置能够在RF放电期间以0.1°C的温度分辨率和最小的温度进行多点实时温度监测1毫米的空间分辨率。在这些实验中,通过在受检器官内部注入温度为37°C的水来进行血液灌注。该解决方案旨在模拟临床实践中进行的射频热消融(RFTA)治疗。