首页> 外文会议>International Instrumentation and Measurement Technology Conference >Evaluation of blood pressure by analyzing the fluctuation frequency of oxygen saturation in palmar digital vein
【24h】

Evaluation of blood pressure by analyzing the fluctuation frequency of oxygen saturation in palmar digital vein

机译:通过分析掌指静脉血氧饱和度的波动频率来评估血压

获取原文

摘要

This paper presents a non-contact measurement method for the evaluation of human blood pressure. A reflection type optical imaging system was developed to interactively capture the human palmar back images with illumination of dual light sources. By a continuous measurement, the variation of oxygen saturation in palmar digital vein could be monitored. The fluctuation frequency of the oxygen saturation was also obtained by FFT algorithm. Through the comparison with electronic sphygmomanometer, the study succeeded in establishing a relationship between the natural blood pressure of human and venous fluctuation frequency. Also, the influence of external pressure for blood vessels on fluctuation frequency was studied. The results show that the fluctuation frequency of oxygen saturation linearly increases with the natural blood pressure or external pressure decreasing. The proposed innovations represent a solution for remote medical diagnosis technology with low risk of infection and high biocompatibility.
机译:本文提出了一种非接触式的人体血压评估方法。开发了一种反射型光学成像系统,以在双光源的照明下交互式捕获人的手掌背部图像。通过连续测量,可以监测掌指静脉中氧饱和度的变化。还通过FFT算法获得了氧饱和度的波动频率。通过与电子血压计的比较,该研究成功地建立了人的自然血压与静脉波动频率之间的关系。此外,研究了外部压力对血管的波动频率的影响。结果表明,氧饱和度的波动频率随着自然血压或外压的降低而线性增加。提出的创新代表了一种具有低感染风险和高生物相容性的远程医学诊断技术的解决方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号