首页> 外文会议>IEEE International Symposium on Medical Measurements and Applications >Focused ultrasound temperature effect in tissue-mimicking material and sheep liver
【24h】

Focused ultrasound temperature effect in tissue-mimicking material and sheep liver

机译:超声对组织模仿材料和绵羊肝脏的聚焦作用

获取原文
获取外文期刊封面目录资料

摘要

The temperature distribution produced by a HIFU (High Intensity Focused Ultrasound) transducer has been investigated by a set up using T type thermocouple temperature sensors in the automated scanning system. The temperature change and the temperature distribution due to applied ultrasound power have been characterized. The temperature measurements were carried out in a tissue-mimicking material (TMM) that has characteristics very similar to human tissue and in ex vivo tissue. A temperature measurement set up in a tissue phantom has been constructed at TÜBİTAK UME (The Scientific and Technological Research Council of Turkey, the National Metrology Institute) Ultrasound laboratory. Burning and removing the tissue by high power ultrasound have been showed.
机译:HIFU(高强度聚焦超声)换能器产生的温度分布已通过在自动扫描系统中使用T型热电偶温度传感器的设置进行了研究。已经表征了由于施加的超声功率引起的温度变化和温度分布。温度测量是在组织模拟材料(TMM)中进行的,该材料具有与人体组织和离体组织非常相似的特征。在TÜBİTAKUME(土耳其科学技术研究委员会,国家计量学会)超声实验室中,已建立了在人体模型中建立的温度测量装置。已经显示了通过高功率超声燃烧和去除组织。