An Automatic Sensing Test procedure for Implantable Cardioverter Defibrillators

机译:植入式心脏复律除颤器的自动感应测试程序

获取原文

摘要

An Automatic Sensing Test procedure (AST) for evaluating the sensing performances of Implantable Cardioverter Defibrillators (ICDs) is presented in this paper. AST is based on the Sensitivity Test described by the Standard CEI EN 45502-2-1 for pacemakers and consists of determining the ICD sensing thresholds at different heartbeat frequencies. Consequently, an extensive evaluation of ICD sensing performances can be achieved. AST has been implemented by means of virtual instruments developed in the Labview environment and has been validated experimentally through several tests. These have been performed inside an anechoic RF chamber. Such an environment guarantees proper shielding against external electromagnetic interferences, thus ensuring reliability and repeatability of the tests. The discussion of the results highlights the usefulness of the proposed procedure.
机译:本文提出了一种自动感应测试程序(AST),用于评估植入式心脏复律除颤器(ICD)的感应性能。 AST基于标准CEI EN 45502-2-1针对起搏器描述的灵敏度测试,包括确定不同心跳频率下的ICD感应阈值。因此,可以实现对ICD感测性能的广泛评估。 AST已通过在Labview环境中开发的虚拟仪器实现,并已通过多项测试进行了实验验证。这些是在无回波射频室内进行的。这样的环境可确保适当屏蔽外部电磁干扰,从而确保测试的可靠性和可重复性。对结果的讨论突出了所建议程序的有用性。