首页> 外文会议>Conference on Medical Imaging: Ultrasonic Imaging and Tomography;Society of Photo-Optical Instrumentation Engineers >Limited view reconstructions with transmission ultrasound tomography – clinical implications and quantitative accuracy
【24h】

Limited view reconstructions with transmission ultrasound tomography – clinical implications and quantitative accuracy

机译:透射超声层析成像的有限视野重建–临床意义和定量准确性

获取原文

摘要

The ability to conduct ‘partial field of view’ scans in QT ultrasound transmission ultrasound is investigated. Thestandard tomographic data acquisition is typically conducted in a full 360 degree aperture, which may limit thepossibility of real time intervention and/or patient positioning for medical imaging. Transmission ultrasound has manyadvantages over other imaging modalities such as, it does not emit ionizing radiation, does not require contrast agent,etc. A partial field of view data acquisition in this context is therefore attractive. Three scenarios are investigated:breast, whole body, and orthopedic imaging. A full suite of 180 views at 2 degree intervals and 192 mm in the verticaldirection was collected from the QT ultrasound™ scanner in the orthopedic and whole body scenarios. The verticalextent of the breast image varied with breast size. Subsequent reconstructions were carried out with a subset of theviews incorporated. The contiguous sector of missing angles varied from 8, 16, 32, 40, 60 and 92 degrees. Thedifference between the corresponding images was quantitatively and qualitatively analyzed and compared. It was foundthat a large lacuna (gap) of contiguous data did not significantly (clinically) degrade image quality, and the quantitativevalues, where relevant. The open acquisition scenario allows us to carry out medical intervention as well as potentiallydecrease patient anxiety. The quantitative nature of the degradation is noted and correlated with the missing sectorangles for clinical scenarios, and the implications discussed.
机译:研究了在QT超声传输超声中进行“部分视场”扫描的能力。这 标准断层扫描数据采集通常在完整的360度光圈中进行,这可能会限制 实时干预和/或患者定位以进行医学成像的可能性。传输超声波有很多 与其他成像方式相比的优势,例如,它不发射电离辐射,不需要造影剂, 因此,在这种情况下的部分视场数据采集是有吸引力的。研究了三种情况: 乳房,全身和骨科影像学检查。一整套180度的视角,每2度间隔一次,垂直方向为192毫米 方向是从QT Ultrasonic扫描仪在整形外科和全身情况中收集的。垂直 乳房图像的范围随乳房大小而变化。随后的重建工作是利用 纳入意见。缺失角的连续扇区范围为8、16、32、40、60和92度。这 定量和定性分析和比较相应图像之间的差异。它被找到了 大量连续数据的空白(间隙)并不会(在临床上)显着降低图像质量, 值(如果相关)。开放获取方案使我们能够进行医疗干预以及潜在的 减轻患者的焦虑感。记录了退化的数量性质,并将其与缺失的部门相关联 临床场景的角度,以及所讨论的含义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号