首页> 外文会议>Conference on Precision Electromagnetic Measurements >Characterization of Cable Effects on a Reference Lightning Impulse Voltage Divider
【24h】

Characterization of Cable Effects on a Reference Lightning Impulse Voltage Divider

机译:参考雷电脉冲分压器上电缆效应的表征

获取原文

摘要

Effect of measurement cables on impulse voltage measurements has been widely ignored by the high voltage community. During the last years the cable length effects have been reported but the results have not been fully consistent. This paper will characterize cable effects on a resistive impulse voltage divider using three different test methods. Used methods are step response analysis using convolution, low-voltage analysis with an impulse calibrator, and high-voltage impulse analysis with an impulse generator. Results show that with reasonable cable lengths less than 25 meters, the time parameter errors increase almost linearly with the cable length. In addition, the results show that the applied methods are comparable ways to determine the overall performance of an impulse voltage measuring system, including the cable effect.
机译:高电压社区已广泛忽略了测量电缆对脉冲电压测量的影响。在过去的几年中,已经报道了电缆长度的影响,但是结果并没有完全一致。本文将使用三种不同的测试方法来表征电缆对电阻式脉冲分压器的影响。使用的方法是使用卷积的阶跃响应分析,使用脉冲校准器的低压分析以及使用脉冲发生器的高压脉冲分析。结果表明,在合理的电缆长度小于25米的情况下,时间参数误差几乎随电缆长度线性增加。此外,结果表明,所应用的方法是确定脉冲电压测量系统总体性能(包括电缆效应)的可比方法。