首页> 外文会议>Design, Automation and Test in Europe Conference and Exhibition >A Non-invasive Wearable Bioimpedance System to Wirelessly Monitor Bladder Filling
【24h】

A Non-invasive Wearable Bioimpedance System to Wirelessly Monitor Bladder Filling

机译:无线监测膀胱充盈的非侵入性可穿戴生物阻抗系统

获取原文

摘要

Monitoring of renal function can be crucial for patients in acute care settings. Commonly during postsurgical surveillance, urinary catheters are employed to assess the urine output accurately. However, as with any external device inserted into the body, the use of these catheters carries a significant risk of infection. In this paper, we present a non-invasive method to measure the fill rate of the bladder, and thus rate of renal clearance, via an external bioimpedance sensor system to avoid the use of urinary catheters, thereby eliminating the risk of infections and improving patient comfort. We design and propose a 4-electrode front-end and the whole wearable and wireless system with low power and accuracy in mind. The results demonstrate the accuracy of the sensors and low power consumption of only 80μW with a duty cycling of 1 acquisition every 5 minutes, which makes this battery-operated wearable device a long-term monitor system.
机译:肾功能的监测对于急性护理环境中的患者至关重要。通常在术后监视期间,使用导尿管来准确评估尿量。但是,如同将任何外部设备插入体内一样,使用这些导管会带来很大的感染风险。在本文中,我们提出了一种非侵入性方法,可通过外部生物阻抗传感器系统来测量膀胱的填充率,从而测量肾脏的清除率,从而避免使用导尿管,从而消除了感染的风险并改善了患者舒适。我们设计并提出一种4电极前端以及整个可穿戴和无线系统,同时考虑到低功耗和准确性。结果表明,传感器的精度和低功耗仅为80μW,占空比每5分钟采集1次,这使这种电池供电的可穿戴设备成为长期监控系统。