We hypothesize that the combination of ultrasound Short-Lag Spatial Coherence (SLSC) imaging and Multi-Line Transmission (MLT) can provide enhanced visibility of highly-coherent targets within soft tissues, improving the frame-rate at the same time. Images of a biopsy needle inserted within a lab-made agar/gelatin phantom and an ex vivo bovine meat sample were acquired after implementing single-line transmission (SLT) and MLT with an increasing number of simultaneously transmitted beams. The received signals were then beamformed with standard Delay and Sum (DAS) and SLSC algorithms. Results show that MLT SLSC images have darker background than SLT SLSC and DAS B-mode images, which is due to the rapid signal decorrelation at short lags caused by inter-beam interferences. This effect in turn provides increased needle contrast with MLT SLSC imaging, particularly as the number of multiple beams increase, which is promising for a more accurate biopsy needle detection.
机译:我们假设超声短滞后空间相干性(SLSC)成像和多线传输(MLT)的组合可以提供软组织内高度相干靶标的增强的可见性,同时提高帧速率。在实现单线透射(SLT)和MLT之后,在实施单线传输(SLT)和MLT之后,在实验室制造的琼脂/明胶模型和离体牛肉样品中的图像进行了图像的图像。然后通过标准延迟和总和(DAS)和SLSC算法进行波束形成所接收的信号。结果表明,MLT SLSC图像具有比SLT SLSC和DAS B模式图像更暗的背景,这是由于由束间干扰引起的短滞后的快速信号去相关性。这种效果又提供了与MLT SLSC成像的增加的针对对比度,特别是随着多个光束的数量增加,这对更准确的活检针检测具有很大的感知。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号