首页> 外文会议>IEEE International Ultrasonics Symposium >Investigating a CMUT's Ability to Achieve Non-linear Contrast Enhancement
【24h】

Investigating a CMUT's Ability to Achieve Non-linear Contrast Enhancement

机译:研究CMUT实现非线性对比度增强的能力

获取原文

摘要

Capacitive micromachined ultrasonic transducers (CMUTs) are inherently non-linear and often believed to be unsuitable for non-linear contrast-enhanced ultrasound (CEUS) imaging. The ability of CMUTs to perform contrast-enhanced ultrasound imaging has been investigated by experimentally evaluating the contrast-to-noise ratio (CTR) and its variation with emitted pressure. This has been evaluated using amplitude modulation, pulse inversion and B-mode imaging sequences. The study shows that the CTR of the AM-mode was 7.5 dB higher than the B-mode on average, while the PI-mode is 5.9 dB lower than the B-mode. This indicates that contrast-enhancement can be achieved with a CMUT when a suitable image sequence is chosen.
机译:电容式微加工超声换能器(CMUT)本质上是非线性的,通常被认为不适用于非线性对比增强超声(CEUS)成像。通过实验评估对比噪声比(CTR)及其随发射压力的变化,已经研究了CMUT执行对比增强型超声成像的能力。已使用幅度调制,脉冲反转和B模式成像序列对此进行了评估。研究表明,AM模式的CTR平均比B模式高7.5 dB,而PI模式比B模式低5.9 dB。这表明,对比度增强可以与CMUT选择合适的图像序列时可实现。