ADVANCED SURFACE SYSTEM FOR CENTRAL VENOUS CATHETER INSERTION TRAINING

机译:中央静脉导管插入训练的先进表面系统

获取原文

摘要

An advanced surface for Central Venous Catheterization (CVC) training and evaluation was designed using sensor ization techniques, including the use of a hall effect sensor array to measure the insertion depth of a catheter. The sensor array was tested for accuracy in both static and dynamic scenarios, and was found to be sufficiently accurate; measuring position with an accuracy of ±1.1 mm on average. The highest deviations in measured positions were located at the extreme ends of the array where calculations rely on only a single sensor. The maximum deviation in measured position was found to be 3.5 mm. This low-cost system of catheter measurement has the potential to improve feedback and assessment of CVC training.
机译:设计用于中心静脉导管(CVC)训练和评估的先进表面,包括使用传感器型技术,包括使用霍尔效应传感器阵列来测量导管的插入深度。 在静态和动态场景中测试传感器阵列以精确度,并被发现足够准确; 测量位置平均精度为±1.1 mm。 测量位置的最高偏差位于阵列的极端端部,其中计算依赖于单个传感器。 发现测量位置的最大偏差为3.5mm。 这种低成本的导管测量系统有可能改善CVC训练的反馈和评估。