Image Quality assessment for Awake Animal Brain PET

机译:清醒动物脑PET的图像质量评估

获取原文

摘要

A restrictive issue for small-animal positron emission tomography (PET) imaging is the fact that animals are either anaesthetized or physically restrained to prevent them from moving during the imaging procedure. Tracking the motion of the animal, based on e.g. radioactive point sources which are attached to the head of the animal, allows to scan awake animals. The use of radioactive point sources for motion tracking has the benefit that it can be used in scanners with small bore size or occluded views, is minimally invasive and does not require additional precise calibration between the PET camera and an external tracking device. The aim of this study was to assess the image quality and quantitative accuracy for PET scans of moving objects, with motion correction based on radioactive point sources, following the procedures of the NEMA NU 4-2008 standard. We found that even extensive motion of an object, including rotation and turning front to back, affects image quality as well as quantitation only marginally compared to the motion free scan.
机译:小动物正电子发射断层扫描(PET)成像的一个局限性问题是对动物进行麻醉或对其进行物理约束以防止它们在成像过程中移动的事实。跟踪动物的运动,例如基于附着在动物头部的放射性点源可以扫描清醒的动物。使用放射性点源进行运动跟踪的好处是,它可以用于孔径较小或被遮挡的扫描仪中,具有最小的侵入性,并且不需要在PET摄像机和外部跟踪设备之间进行额外的精确校准。这项研究的目的是根据NEMA NU 4-2008标准的程序,通过基于放射性点源的运动校正来评估运动对象的PET扫描的图像质量和定量准确性。我们发现,与自由运动扫描相比,即使对象的广泛运动(包括旋转和从前向后旋转)也仅对图像质量和定量产生了很小的影响。