首页> 外文会议>IEEE Nuclear Science Symposium;Medical Imaging Conference >Novel volume based approach to estimate contrast recovery for NEMA image quality phantom
【24h】

Novel volume based approach to estimate contrast recovery for NEMA image quality phantom

机译:基于体积的新颖方法可估算NEMA图像质量体模的对比度恢复

获取原文

摘要

In this paper a new recipe for precise measurement of contrast recovery in NEMA image quality phantom for PET scanners is proposed. To accomplish this, we developed flexible numerical tool in three dimensional Cartesian to compute contrast recovery inside volume of interest while accounting of partial volume effects at the boundary of sphere. Also a method has been developed to find a location of centroid for each hot sphere inside NEMA IQ phantom automatically. We compared our proposed method (3D) to current NEMA NU2 2018 standard (2D) used to estimate contrast recovery.
机译:在本文中,提出了一种用于在PET扫描仪的NEMA图像质量模型中精确测量对比度恢复的新方法。为此,我们在三维笛卡尔中开发了灵活的数值工具,以计算感兴趣体积内的对比度恢复,同时考虑了球体边界处的部分体积效应。还开发了一种方法,可以自动为NEMA IQ体模内的每个热球找到质心的位置。我们将建议的方法(3D)与用于估计对比度恢复的当前NEMA NU2 2018标准(2D)进行了比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号