【24h】

Shadow Attacks on MEDA Biochips

机译:对MEDA生物芯片的影子攻击

获取原文

摘要

The Micro-electrode-dot-array (MEDA) is a next-generation digital microfluidic biochip (DMFB) platform that supports fine-grained control and real-time sensing of droplet movements. These capabilities permit continuous monitoring and checkpoint-based validation of assay execution on MEDA. This paper presents a class of “shadow attacks” that abuse the timing slack in the assay execution. State-of-the-art checkpoint-based validation techniques cannot expose the shadow operations. We develop a defense that introduces extra checkpoints in the assay execution at time instances when the assay is prone to shadow attacks. Experiments confirm the effectiveness and practicality of the defense.
机译:微电极点阵列(MEDA)是下一代数字微流生物芯片(DMFB)平台,支持精细控制和液滴运动的实时感测。这些功能允许对MEDA上的测定执行进行连续监控和基于检查点的验证。本文介绍了一类“影子攻击”,它们在检测执行过程中滥用了时间上的懈怠。最新的基于检查点的验证技术无法公开影子操作。我们开发了一种防御措施,可在化验容易受到影子攻击的情况下,在化验执行中引入额外的检查点。实验证实了防御的有效性和实用性。