摘要

Quality of chest compressions (CCs) is key for patient's survival to cardiac arrest. Incorporating real-time feedback on CC rate in automated external defibrillators (AEDs) could enhance CC quality. The aim of the study was to design a method for detecting CC pauses and computing CC rate using the transthoracic impedance (TI) signal acquired from the AED defibrillation pads. We collected 35 AED recordings from a database property of Emergentziak-Osakidetza (Basque Country, Spain). Using the ECG and TI signals, a CC pause was annotated when CCs were interrupted for more than 3 s. The algorithm estimates the CC rate from the autocorrelation of 2-s consecutive TI intervals. The algorithm reported a sensitivity (Se) and a positive predictive value (PPV) in the detection of CC pauses of 97.64%/92.48%. Reliable CC rates were obtained with a Se/PPV of 99.03%/98.61%. Unreliable CC rates were reported only in 0.9% of the cases. A method based on the autocorrelation of the TI signal allows reliable detection of CC pauses and estimation of CC rates, enhancing current AED functionality.
机译:胸外按压(CCs)的质量是患者心脏骤停生存的关键。在自动体外除颤器(AED)中结合CC速率的实时反馈可以提高CC的质量。该研究的目的是设计一种方法,该方法使用从AED除颤垫获取的经胸阻抗(TI)信号来检测CC暂停并计算CC速率。我们从Emergentziak-Osakidetza(西班牙巴斯克地区)的数据库属性中收集了35份AED录音。使用ECG和TI信号,当CC中断超过3 s时,便会注释CC暂停。该算法从2 s个连续TI间隔的自相关估计CC速率。该算法在检测CC暂停时报告的灵敏度(Se)和阳性预测值(PPV)为97.64%/ 92.48%。 Se / PPV为99.03%/ 98.61%,可获得可靠的CC率。仅在0.9%的案例中报告了不可靠的CC率。一种基于TI信号自相关的方法,可以可靠地检测CC暂停并估算CC速率,从而增强当前的AED功能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号