首页> 外文会议>Conference on physics of medical imaging >Characterization of Various Tissue Mimicking Materials for Medical Ultrasound Imaging
【24h】

Characterization of Various Tissue Mimicking Materials for Medical Ultrasound Imaging

机译:用于医学超声成像的各种组织模拟材料的表征

获取原文

摘要

Tissue mimicking materials are physical constructs exhibiting certain desired properties, which are used in machine calibration, medical imaging research, surgical planning, training, and simulation. For medical ultrasound, those specific properties include acoustic propagation speed and attenuation coefficient over the diagnostic frequency range. We investigated the acoustic characteristics of polyvinyl chloride (PVC) plastisol, polydimethyl-siloxane (PDMS), and isopropanol using a time-of-flight technique, where a pulse was passed through a sample of known thickness contained in a water bath. The propagation speed in PVC is approximately 1400 m s~(-1) depending on the exact chemical composition, with the attenuation coefficient ranging from 0.35 dB cm~(-1) at 1 MHz to 10.57 dB cm~(-1) at 9 MHz. The propagation speed in PDMS is in the range of 1100 m s~(-1), with an attenuation coefficient of 1.28 dB cm~(-1) at 1 MHz to 21.22 dB cm~(-1) at 9 MHz. At room temperature (22°C), a mixture of water-isopropanol (7.25% isopropanol by volume) exhibits a propagation speed of 1540ms~(-1), making it an excellent and inexpensive tissue-mimicking liquid for medical ultrasound imaging.
机译:模拟组织的材料是具有某些所需特性的物理构造,可用于机器校准,医学成像研究,手术计划,培训和模拟。对于医学超声,那些特定的属性包括在诊断频率范围内的声传播速度和衰减系数。我们使用飞行时间技术研究了聚氯乙烯(PVC)增塑溶胶,聚二甲基硅氧烷(PDMS)和异丙醇的声学特性,其中脉冲通过了水浴中已知厚度的样品。根据确切的化学成分,PVC中的传播速度约为1400 ms〜(-1),衰减系数的范围从1 MHz时的0.35 dB cm〜(-1)到9 MHz时的10.57 dB cm〜(-1) 。 PDMS中的传播速度在1100 m s〜(-1)范围内,在1 MHz时的衰减系数为1.28 dB cm〜(-1),在9 MHz时的衰减系数为21.22 dB cm〜(-1)。在室温(22℃)下,水-异丙醇(体积比为7.25%异丙醇)的混合物的传播速度为1540ms〜(-1),使其成为用于医学超声成像的出色且廉价的组织模拟液。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号