Front-end electronics for cable reduction in Intracardiac Echocardiography (ICE) catheters

机译:用于减少心内超声(ICE)导管电缆的前端电子设备

获取原文

摘要

3-D imaging ICE catheters with large element counts present design challenges in achieving simultaneous data readout from all elements while significantly reducing cable count for a small catheter diameter. Current approaches such as microbeamformer techniques tend to rely on area and power hungry circuits, making them undesirable for ICE catheters. In this paper, a system which uses are an efficient real-time programmable on-chip transmit (TX) beamformer circuitry to reduce the cable count on the TX side and analog 8:1 Time Division Multiplexing (TDM) with Direct Digital Demodulation (DDD) to reduce the cable count on the receive (RX) side is presented.
机译:具有大元素的3-D成像冰导管对所有元素的同时数据读出进行了目前的设计挑战,同时显着减少了小导管直径的电缆计数。 MicrobeamFormer技术的电流方法倾向于依靠区域和功率饥饿的电路,使其对冰导管不希望。在本文中,使用的系统是一个有效的实时可编程片上发射(TX)波束形成器电路,以减少TX侧的电缆计数和具有直接数字解调的模拟8:1时分复用(TDM)(DDD )提出了减少接收(RX)侧的电缆计数。