Front-end electronics for cable reduction in Intracardiac Echocardiography (ICE) catheters

机译:用于心内超声心动图(ICE)导管中电缆减少的前端电子设备

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

3-D imaging ICE catheters with large element counts present design challenges in achieving simultaneous data readout from all elements while significantly reducing cable count for a small catheter diameter. Current approaches such as microbeamformer techniques tend to rely on area and power hungry circuits, making them undesirable for ICE catheters. In this paper, a system which uses are an efficient real-time programmable on-chip transmit (TX) beamformer circuitry to reduce the cable count on the TX side and analog 8/1 Time Division Multiplexing (TDM) with Direct Digital Demodulation (DDD) to reduce the cable count on the receive (RX) side is presented.
机译:具有大量元素的3-D成像ICE导管在实现从所有元素同时读取数据的同时,还显着减少了导管直径较小的电缆数量时所面临的设计挑战。诸如微波束形成器技术之类的当前方法趋向于依赖于面积和功率消耗电路,这使得它们对于ICE导管而言是不期望的。本文中的系统使用高效的实时可编程片上发送(TX)波束形成器电路来减少TX端的电缆数量,并采用直接数字解调(DDD)技术进行模拟8/1时分多路复用(TDM) ),以减少接收(RX)端的电缆数量。