首页> 外文会议>IEEE biomedical circuits and systems conference >Experimental measurements of multi-carrier based Doppler heartrate tracker
【24h】

Experimental measurements of multi-carrier based Doppler heartrate tracker

机译:基于多载波的多普勒心率追踪器的实验测量

获取原文

摘要

This study presents experimental results of multi-carrier based Doppler radar signal processing for vital signs tracking. When small motion-transients exist, it is expected that a disturbance at a carrier frequency is different from other carrier frequencies. The multi-carrier based Doppler radar shows the possibility to overcome the motion-transient effect in some situations. Six tests are performed to measure multi-carrier Doppler signals. Their heartrate tracking results show that the mean values of the tracked heartrates are within plus/minus 3% of the ground truth R-R interval frequency of their electrocardiogram measurements. It is also shown that the variance values of the tracked heartrates are all within 0.0062 Hz. The proposed signal processing scheme seems to have potential to track vital signs while small motion-transients exist.
机译:这项研究提出了基于多载波的多普勒雷达信号处理生命体征跟踪的实验结果。当存在小的运动瞬变时,可以预期在载波频率上的干扰不同于其他载波频率。基于多载波的多普勒雷达显示了在某些情况下克服瞬态运动影响的可能性。进行了六个测试以测量多载波多普勒信号。他们的心率跟踪结果表明,所跟踪的心率的平均值在其心电图测量结果的地面真实R-R间隔频率的正负3%之内。还显示跟踪的心率的方差值都在0.0062 Hz之内。所提出的信号处理方案似乎有可能在存在较小的运动瞬变时跟踪生命体征。