Telemetric system for monitoring of endoleak in abdominal aorta aneurysm using multiple pressure sensors integrated on a stent graft

机译:使用集成在覆膜支架上的多个压力传感器的遥测系统,用于监测腹主动脉瘤内漏

获取原文

摘要

An abdominal aortic aneurysm (AAA) is the dilation of the abdominal aorta due to weakening of the arterial wall. Endovascular aneurysm repair (EVAR) using an implantable stent graft is the commonly used method for treatment of AAA. An endoleak, which is the leakage of blood around the stent graft, is the complication after EVAR creating trauma to the weakened arterial walls of the aneurysm leading to further enlargement and risk of rupture. This paper details the design, implementation and testing of a wirelessly powered implant device containing multiple pressure sensors to be mounted on a stent graft for detecting endoleak. The tested implant device can be powered for distance of 20 cm using an external RF power source of 2.65 W. A stent prototype with sixteen pressure sensors is fabricated and tested using an in-vitro chamber. The resolution and directionality of the pressure sensors in the stent prototype are measured. Different wire bonding glob top encapsulations are evaluated for mechanical stress induced during stent delivery.
机译:腹主动脉瘤(AAA)是由于动脉壁变弱而引起的腹主动脉扩张。使用可植入支架移植物的血管内动脉瘤修复(EVAR)是治疗AAA的常用方法。内漏是支架移植物周围的血液泄漏,是EVAR对动脉瘤变弱的动脉壁造成创伤后导致进一步扩大和破裂风险的并发症。本文详细介绍了包含多个压力传感器的无线植入设备的设计,实施和测试,该传感器将安装在用于检测内漏的支架移植物上。可以使用2.65 W的外部RF电源为经过测试的植入设备供电20厘米的距离。使用十六个压力传感器制造支架原型,并使用体外腔进行测试。测量了支架原型中压力传感器的分辨率和方向性。针对支架递送期间引起的机械应力,评估了不同的引线键合球形顶部封装。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号