A study of sparse detector designs with interpolation for multi-slice spiral CT

机译:多层螺旋CT插值稀疏检测器设计的研究

获取原文

摘要

In multi-slice spiral CT, z-sampling has a great effect on image quality, dose and speed. In our work, we explore using sparse detectors of various designs in multi-slice spiral CT. These designs have fewer detector elements as compared to a regular detector. Or alternatively, they have the same number of detectors but provide more axial coverage for greater scanning speed of longer objects. The sparser detector elements also affect the reconstruction in a similar manner as z-sampling does. To account for the missing detector elements we interpolate the acquired data to simulate a full detector. We apply both bilinear interpolation and directional interpolation and compare them. We also report the quality of the reconstruction and scan speed for each detector pattern.
机译:在多层螺旋CT中,z采样对图像质量,剂量和速度有很大影响。在我们的工作中,我们探索在多层螺旋CT中使用各种设计的稀疏检测器。与常规检测器相比,这些设计具有更少的检测器元件。或者,它们具有相同数量的检测器,但提供了更大的轴向覆盖范围,以实现更长的物体更快的扫描速度。稀疏检测器元素还以与z采样类似的方式影响重建。为了说明缺少的检测器元素,我们对获取的数据进行插值以模拟完整的检测器。我们同时应用了双线性插值和方向插值,并对其进行了比较。我们还报告了每种检测器模式的重建质量和扫描速度。