The objective of this study is to assess the clinical performance of a novel bi-directional axial transmission (BDAT) ultrasound prototype. The speed of sound of the first arriving signal (V) was measured at the radius (single site) and at the tibia (two sites) of 51 post-menopausal females. An optimized V measurement protocol was introduced to reduce the clinical measurement time (from 30 to 10 minutes) by maintaining the robustness of the former method. The optimized protocol was efficient and showed excellent discrimination performance: the highest areas under the ROC curves (AUC) were found for V radius (OR: 3.54; AUC: 0.81) and V tibia medial surface (MS) (OR: 4.10; AUC: 0.81). Dual energy x-ray absorptiometry (DXA) at the lumbar spine exhibited similar fracture discrimination power (OR: 1.89; AUC: 0.81) compared to the ultrasonic method.
机译:这项研究的目的是评估新型双向轴向传输(BDAT)超声原型的临床性能。在51位绝经后女性的半径(单个部位)和胫骨(两个部位)处测量第一个到达信号(V)的声速。通过保持前一种方法的鲁棒性,引入了优化的V测量方案以减少临床测量时间(从30到10分钟)。优化后的协议有效且显示出出色的判别性能:发现ROC曲线(AUC)下的最高区域为V半径(OR:3.54; AUC:0.81)和V胫骨内侧表面(MS)(OR:4.10; AUC: 0.81)。与超声方法相比,腰椎的双能X线骨密度仪(DXA)表现出相似的骨折鉴别能力(OR:1.89; AUC:0.81)。