This paper presents a custom-designed single-cable readout IC for the reception of the broadband photoacoustic (PA) signal in intravascular photoacoustic (IVPA) imaging. The readout IC is intended for direct integration behind a broadband polyvinylidene fluoride (PVDF) transducer in an IVPA catheter tip to match the impedance between the small PVDF element and the connecting cable. The capability of the readout IC to work with a single cable that combines the output signal and the power supply ensures the mechanical flexibility of the IVPA catheter. Electrical measurements show that the readout IC provides a flat frequency response from 1 MHz to 20 MHz with a 6 mA external current supply. The acoustical measurements involving the readout IC and the PVDF transducer demonstrate a 60 dB dynamic range, a sensitivity of 3.8 μV/Pa at 2.25 MHz, and a broad receiving bandwidth from 2 MHz to 15 MHz.
机译:本文提出了一种定制设计的单电缆读出IC,用于在血管内光声(IVPA)成像中接收宽带光声(PA)信号。读出IC旨在直接集成在IVPA导管末端的宽带聚偏二氟乙烯(PVDF)传感器后面,以匹配小型PVDF元件和连接电缆之间的阻抗。读出IC可以使用结合了输出信号和电源的单根电缆来工作,从而确保了IVPA导管的机械灵活性。电气测量表明,读出IC具有6 mA外部电源,可提供从1 MHz到20 MHz的平坦频率响应。涉及读出IC和PVDF传感器的声学测量表明,动态范围为60 dB,2.25 MHz时的灵敏度为3.8μV/ Pa,接收带宽为2 MHz至15 MHz。