Application of HDMI? Cables as an MRI Compatible Single Cable Solution for Readout and Power Supply of SiPM Based PET Detectors

机译:HDMI的应用?电缆作为MRI兼容的单电缆解决方案,用于基于SIPM的PET探测器的读数和电源

获取原文

摘要

All PET detectors require cabling to supply power and transmit data, whether in analog or digital form. In this work we examine using High-Definition Multimedia Interface (HDMI~?) cables as a single cable solution to both supply power and transmit analog signals for a SiPM based PET detector that will be used in a hybrid PET/MRI imaging system. HDMI cables are designed for digital audio/video transmission, with each cable having four pairs of shielded 100 ohm differential signal lines, each rated for 3.4 Gbit/s, along with 7 additional signal lines and a braided shield around the entire cable. While designed as a digital signal cable, the lines are rated for 40V DC and 0.5A current. To test using HDMI? cables for application in PET, a silicon photomultiplier (SiPM) PET detector module with a single HDMI~? Type C ('mini) receptacle was built using a SensL SPMArray4 coupled to a dual-layer offset LYSO crystal array 1.67 mm crystal pitch. The 16 detector outputs were multiplexed to four signals using a simple resistor charge division network and driven using Analog Devices AD8132 op-amps. Detector bias of 30.2V and +/- 5V power for the op-amps was supplied through a 25' length HDMI? cable connected through a custom receiver card. The signals were processed using NIM electronics and digitized with a PC based analog to digital converter (ADC) card. The PET detector had an average energy resolution of 14.3% full-width at half-maximum (FWHM) and an average timing resolution of 2.50 ns. The performance of the PET detector was unaffected when operated inside a 7T MRI system. The MR images showed degradation in signal to noise ratio (SNR) of between 7 and 23%, which is believe to be due to coupling between the RF coil and the PET detector shielding. The results of this work suggest that HDMI~? cables with the Type C connector are a viable, cost effective and robust single cable solution for both analog signal and power/bias transmission for an MRI compatible PET insert system.
机译:无论是模拟还是数字形式,所有PET探测器都需要电缆供电和传输数据。在这项工作中,我们将使用高清多媒体接口(HDMI〜Δ)电缆作为单个电缆解决方案,用于电源功率和传输模拟信号,用于基于SIPM的PET检测器,该宠物检测器将用于混合宠物/ MRI成像系统。 HDMI电缆专为数字音频/视频传输而设计,每个电缆具有四对屏蔽100欧姆差分信号线,每个电缆为3.4 Gbit / s等级,以及7个附加信号线和整个电缆周围的编织屏蔽。设计为数字信号电缆时,线路额定电流为40V直流和0.5A电流。使用HDMI进行测试?用于PET的电缆,硅光电倍增器(SIPM)PET检测器模块,单个HDMI〜?使用SECSL SPMARRAY4构建C型('mini)插座,耦合到双层偏移Lyso晶体阵列1.67mm晶间距。使用简单的电阻器电荷划分网络将16个检测器输出复用到四个信号,并使用模拟设备AD8132 OP-AMPS驱动。通过25'长度HDMI提供30.2V和OP-AMPS的30.2V和+/- 5V电源的探测器偏差?电缆通过自定义接收器卡连接。使用NIM电子设备处理信号,并使用基于PC的模拟转换器(ADC)卡进行数字化。 PET检测器的平均能量分辨率为14.3%的全宽度为半最大(FWHM),平均定时分辨率为2.50ns。当在7T MRI系统内部操作时,PET检测器的性能不受影响。 MR图像显示出信号到噪声比(SNR)的降解,其中7-23%,这相信是由于RF线圈和PET检测器屏蔽之间的耦合。这项工作的结果表明HDMI〜? C连接器的电缆是用于MRI兼容PET插入系统的模拟信号和电源/偏置传输的可行性,成本效益和强大的单电缆解决方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号