首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >LabPET II, a Novel 64-channel APD-based PET Detector Module with Individual Pixel Readout Achieving Submillimetric Spatial Resolution
【24h】

LabPET II, a Novel 64-channel APD-based PET Detector Module with Individual Pixel Readout Achieving Submillimetric Spatial Resolution

机译:Labpet II,一种新型的64通道APD的PET检测器模块,具有实现淹没空间分辨率的单个像素读数

获取原文

摘要

A new avalanche photodiode (APD) detector module, the LabPET II, was developed to achieve submillimetre spatial resolution for small animal molecular imaging. The module consists of two monolithic APD arrays of 4 × 8 pixels, each with an active area of 1.1 × 1.1 mm~2 at a 1.2 mm pitch. The two APD arrays are mounted in a custom ceramic holder and coupled to an 8 × 8 LYSO scintillator array designed to accommodate one-to-one coupling between individual APDs and crystal pixels. An analog test board adapted from the LabPET processing electronics and consisting of four 16-channel preamplifier ASICs, was designed for testing the LabPET II detector module. The devices have a broad operating range (over 200 V) with breakdown voltage of ~350 V, at which a typical gain well above 200 is reached. Individual APD pixels have a capacitance of 3.7 ± 0.4 pF (including stray capacitance), a typical dark current of 30 ± 11 nA, a dark noise of 0.13 ± 0.03 pA/Hz~(1/2) and an equivalent noise charge (ENC) of 12 e~- rms at a gain of 100. At a standard APD operating bias, a mean energy resolution of 27.5 ± 2.1% was typically obtained with a relative standard deviation of 13.8% in signal amplitude for the 64 individual pixels when irradiated with 511 keV photons. A global timing resolution of 5.0 ± 0.2 ns FWHM was measured with two modules in coincidence. Finally, an intrinsic spatial resolution of 0.82 ± 0.02 mm FWHM (1.54 ± 0.05 mm FWTM) was obtained by sweeping a ~(22)Na point source between two rows of the detector array. By deconvolving the source size and non-collinearity, an expected 0.73 mm intrinsic geometric crystal resolution is obtained. The LabPET II detector module is demonstrating promising characteristics for dedicated small animal PET imaging at submillimetre resolution and, with some further optimization, would be suitable as the building block for a dual-modality combined PET/CT system.
机译:开发了一种新的雪崩光电二极管(APD)检测器模块,Labpet II,实现小动物分子成像的亚倍仪空间分辨率。该模块由两个单片APD阵列为4×8像素组成,每个单片APD阵列,各个有效面积为1.1×1.1mm〜2的间距。两个APD阵列安装在自定义陶瓷支架中,并耦合到8×8 Lyso闪烁体阵列,该闪烁器阵列被设计成适应各个APD和晶体像素之间的一对一耦合。设计从Labpet处理电子设备的模拟测试板,并由四个16通道前置放大器ASIC组成,用于测试Labpet II检测器模块。该器件具有宽的工作范围(超过200V),击穿电压为〜350 V,此处达到典型的增益超过200。单个APD像素的电容为3.7±0.4pf(包括杂散电容),典型的暗电流为30±11 na,暗噪声为0.13±0.03 pa / hz〜(1/2)和等效噪声充电(ENC )在100的增益中的12 e〜RMS。在标准APD操作偏压下,通常在照射时的64个单个像素的信号幅度的相对标准偏差为27.5±2.1%的平均能量分辨率。用511 kev光子。用两种模块巧合,测量全球定时分辨率为5.0±0.2ns fwhm。最后,通过在探测器阵列的两行之间扫描〜(22)NA点源来获得0.82±0.02mm FWHM(1.54±0.05mm FWTM)的固有空间分辨率。通过解构源极尺寸和非共线性,获得预期的0.73mm内部几何晶体分辨率。 Labpet II探测器模块正在展示在亚瑟赛中的专用小型动物宠物成像的有希望的特性,并且随着一些进一步的优化,适合作为双模组合PET / CT系统的构建块。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号