Micropatterned Gaseous Detectors (MPGDs) have been used for several years in single photon counting and energy resolved imaging systems. This kind of detectors presents suitable characteristics for such purposes as: elimination of the electronic noise, good energy resolution and high counting rate capability, absence of dead areas and 2D intrinsic position discrimination capability. A Computed Tomography (CT) system using a Thick COBRA (THCOBRA) with energy resolution capability envisaging small animal imaging is being developed. By using this kind of single photon counting and energy resolved detector it is possible to retain the interaction position information and energy of each detected single photon, allowing the application of Energy Weighting Technique (EWT) and, therefore, improvement of image quality. Cross-sectional images of a Poly(methyl methacrylate) - PMMA - phantom and of a sea snail are presented.
机译:微图案化气体探测器(MPGDS)已在单光子计数和能量分辨成像系统中使用了几年。这种探测器的目的是:消除电子噪声,良好的能量分辨率和高计数率能力,缺乏死区和2D内在位置辨别能力。正在开发使用带有能量分辨率的厚眼镜蛇(Thcobra)的计算机断层扫描(CT)系统,包括设想小动物成像。通过使用这种单光子计数和能量分辨检测器,可以保留每个检测到的单个光子的相互作用位置信息和能量,允许应用能量加权技术(EWT),并因此提高图像质量。提出了聚(甲基丙烯酸甲酯) - PMMA - 幻像和海蜗牛的横截面图像。