首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >Dual-headed SPECT for awake animal brain imaging
【24h】

Dual-headed SPECT for awake animal brain imaging

机译:用于唤醒动物脑成像的双头SPECT

获取原文

摘要

Motion-corrected awake animal imaging is needed for normal-state investigations of models of neurological disease and brain activity. The awake animal brain SPECT/CT system, AwakeSPECT at Johns Hopkins University has in the past used a single gamma camera for imaging. Enhancements have been made by adding a pinhole collimator to the second gamma camera at the opposite side which has been previously equipped parallel hole collimator. Geometry calibration was performed using a custom built quality control phantom containing three Co-57 point sources and applied to the tomographic reconstruction code. Hot-rod phantom scans with Tc-99m were performed to test sensitivity and resolution improvements. The reconstruction results show significant resolution and sensitivity improvements.
机译:正常状态调查神经疾病和脑活动模型需要运动校正的唤醒动物成像。在Johns Hopkins大学的令人醒着的动物脑SPECT / CT系统中,在过去使用单个伽玛相机进行成像。通过在已经配备了平行孔准直器的相对侧向第二伽马摄像机添加到第二伽马摄像机来进行增强。使用包含三个CO-57点源的定制构建质量控制幻像进行几何校准,并应用于断层切断重建码。使用TC-99M的热杆幻影扫描进行了测试灵敏度和分辨率改进。重建结果显示出显着的分辨率和敏感性。

著录项