首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference >Development of a detector based on Silicon Drift Detectors and 2??? lanthanum bromide scintillator for gamma-ray spectroscopy
【24h】

Development of a detector based on Silicon Drift Detectors and 2??? lanthanum bromide scintillator for gamma-ray spectroscopy

机译:基于硅漂移检测器和2 ???的检测器的开发溴化镧闪烁体,用于伽马射线光谱

获取原文

摘要

This work deals with the development of a new gamma detector based on Silicon Drift Detectors (SDDs) to read out large (2" up to 3") LaBr:Ce scintillators for gamma-ray spectroscopy for astronomy applications, an activity supported by ESA. SDDs, characterized by high quantum efficiency and low electronic noise, are good candidates in such applications with a wide energy range of gamma rays to be detected from 150 keV to 15 MeV. The gamma-ray detection system here presented is composed of a 2"×2" LaBr:Ce scintillator coupled with four SDD modules arranged in a 2×2 format (a total of 36 SDDs). An energy resolution of 3.5% FWHM at 662 keV has been achieved with this detection system being read out by custom developed electronics. In this work we will report on the SDD technology used, their X-ray characterization, readout electronics, experimental setups and gamma-ray spectroscopy results.
机译:这项工作涉及基于硅漂移探测器(SDD)的新伽马探测器的开发,读出大(2“最多3”)Labr:CE闪烁器,用于天文学应用的γ射线光谱,由ESA支持的活动。 SDD,以高量子效率和低电子噪声为特征,是在这种应用中的良好候选者,其具有宽的伽马光线,以检测到150keV至15meV。这里提出的伽马射线检测系统由2“×2”Labr:CE闪烁体耦合,耦合有四个SDD模块,以2×2格式(总共36个SDDS)组成。通过定制开发的电子设备读出该检测系统,已经实现了在662keV的3.5%FWHM的能量分辨率。在这项工作中,我们将报告使用的SDD技术,其X射线表征,读出电子设备,实验设置和伽马射线光谱结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号