This paper presents a bioelectrical impedance analysis circuit with ultralow-current source using gate-induced drain-leakage (GIDL) current for biomedical applications. The proposed circuit consists of an ultralow-current reference circuit, a minimal voltage measurement block, a precise current source block, and a digital control logic circuit. The reference circuit generates pico-ampere-order currents based on GIDL current of an n-channel MOSFET. Fabricated in a 0.18μm 1P6M standard CMOS technology, the impedance analysis circuit occupies 0.27mm2 and can measure impedance range from 100Ω to 10MΩ. Experimental results shows that the fabricated current source using GIDL current generates a quite stable ultralow-current of 50pA, 100pA and 200pA, respectively. In addition, the proposed analysis circuit successfully measures the electrode-tissue interface impedance and tissue impedance by functional verification using monolayer and inhomogeneous agarose gel test setup phantom.
机译:本文介绍了使用栅极诱导的漏极电流(GID1)电流的超级电流源的生物电阻抗分析电路,用于生物医学应用。所提出的电路包括超级电流参考电路,最小电压测量块,精确电流源块和数字控制逻辑电路。参考电路基于N沟道MOSFET的GIDL电流生成微微级订单电流。在0.18μm1p6m标准CMOS技术中制造,阻抗分析电路占用0.27mm2,可以测量阻抗范围从100Ω到10mΩ。实验结果表明,使用GID1电流的制造电流源分别产生了50pa,100pa和200pa的相当稳定的超级电流。此外,所提出的分析电路通过使用单层和非均匀琼脂糖凝胶试验模拟模型通过功能验证成功测量电极组织界面阻抗和组织阻抗。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号