首页> 外文会议>IEEE biomedical circuits and systems conference >3.2 mW ultrasonic LSK modulator for uplink communication in deep implanted medical devices
【24h】

3.2 mW ultrasonic LSK modulator for uplink communication in deep implanted medical devices

机译:3.2 mW超声波LSK调制器,用于深度植入医疗设备中的上行链路通信

获取原文

摘要

The wireless body area network (WBAN) includes up to ten sensor nodes and an external control unit to synchronize the wireless power transfer and communication. A battery operated sensor node is designed using discrete components and tested on standard FR4 substrate. The board dimensions are 2.7 × 4.9 cm which fits inside the Kinetra neurostimulator casing provided by Medtronic, wherein a six elements array transducer (5 × 10 mm) is glued. The transponder power consumption is 3.2 mW when transmitting data towards the control unit, and only 49 µW when in sleep mode.
机译:无线人体局域网(WBAN)包括多达十个传感器节点和一个外部控制单元,用于同步无线功率传输和通信。电池供电的传感器节点是使用分立组件设计的,并在标准FR4基板上进行了测试。板的尺寸为2.7×4.9 cm,适合于Medtronic提供的Kinetra神经刺激器外壳,其中粘合了6个元件阵列的换能器(5×10 mm)。向控制单元发送数据时,应答器功耗为3.2 mW,处于睡眠模式时仅为49 µW。