A measurement of blood pressure is a basic diagnostic technique for examining of the cardiovascular system. There are a lot of methods for determination of the blood pressure, but an oscillometric measurement (investigation of oscillometric pulsations) is one of the most frequent methods used in modern electronic tonometers. The results given by oscillometric measurements are very dependent on the cleanness of the signal. The method is very sensitive for example for motion artifacts, which produce significant damage of the signal. For the proper processing of the signal it is very important to detect all artifacts precisely before signal processing. The paper deals with the methods for artifacts detection in oscillometric pulsations signals. The methods based on predicted heart rate detection, heart rate variability (standard deviation method) and comparison of oscillometric pulsations with ECG signal are described in the paper. The section Results includes a comparison of the methods and a discussion of their advantages and disadvantages. The authors conclude that the algorithm based on detection of heart rate variability (standard deviation method) seems as the most suitable algorithm for automated detection of artifacts in the signal of oscillometric pulsations.
机译:血压测量是检查心血管系统的基本诊断技术。有很多确定血压的方法,但是示波测量(示波脉动的研究)是现代电子眼压计中最常用的方法之一。示波法测量的结果在很大程度上取决于信号的清洁度。该方法对于例如运动伪影非常敏感,该运动伪影会产生信号的严重损坏。为了正确处理信号,在信号处理之前准确检测所有伪像非常重要。本文介绍了示波脉动信号中伪迹的检测方法。本文介绍了基于预测心率检测,心率变异性(标准差法)以及示波脉动与ECG信号比较的方法。结果部分包括方法的比较以及它们的优缺点的讨论。作者得出的结论是,基于心率变异性检测的算法(标准差方法)似乎是自动检测示波脉冲信号中伪像的最合适算法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号