首页> 外文会议>IEEE Symposium on Product Compliance Engineering >Why 500VA for hipot testers? A TC66 investigation to explore the necessity of the 500VA hipot tester
【24h】

Why 500VA for hipot testers? A TC66 investigation to explore the necessity of the 500VA hipot tester

机译:为什么耐压测试仪需要500VA? TC66调查以探讨500VA耐压测试仪的必要性

获取原文

摘要

The IEC 61010-1 3rd Edition standard for measurement and control equipment for laboratory use stipulates that a 500VA dielectric withstand tester must be used to test product insulation as part of testing for electrical hazards. This requirement harks back to a time when most withstand tester outputs were unregulated and could cause problems with load regulation. Advancements in withstand tester technology include testers with regulated outputs. This calls into question the need for a 500VA requirement in the IEC 61010-1 3rd Edition standard.
机译:用于实验室的测量和控制设备的IEC 61010-1 3 rd 版标准规定,必须使用500VA绝缘耐压测试仪来测试产品绝缘,以作为电气危险测试的一部分。这项要求可以追溯到大多数耐压测试仪的输出均未调节并可能导致负载调节问题的时期。耐压测试仪技术的进步包括具有稳定输出的测试仪。这引起了对IEC 61010-1 3 rd 版标准中500VA要求的需求的质疑。