首页> 外文会议>IEEE Wireless Power Transfer Conference >Wireless Power Supply System for Left Ventricular Assist Device and Implanted Cardiac Defibrillator
【24h】

Wireless Power Supply System for Left Ventricular Assist Device and Implanted Cardiac Defibrillator

机译:无线电源系统,用于左心室辅助装置和植入心脏除颤器

获取原文

摘要

This study focuses on the development of a new power supply architecture for an implanted cardiac defibrillator (ICD) and a left ventricular assist device (LVAD). In the proposed solution a transcutaneous wireless power transfer technology (WPT) based on magnetic resonance coupling is adopted assuming an on-body transmitting coil and a subcutaneous receiving coil. With this solution, the percutaneous driveline of the LVAD is eliminated, reducing the risk of infection. The WPT receiver is integrated with the ICD. The proposed power supply using a rechargeable ICD battery allows the LVAD to be continuously powered and the ICD battery to be recharged periodically.
机译:本研究侧重于开发植入心脏除颤器(ICD)和左心室辅助装置(LVAD)的新电源架构。 在所提出的解决方案中,假设体上透射线圈和皮下接收线圈采用基于磁共振耦合的经皮无线电力传递技术(WPT)。 通过这种解决方案,消除了LVAD的经皮传动系,降低了感染的风险。 WPT接收器与ICD集成。 使用可充电ICD电池的提出电源允许LVAD连续供电,并且定期再充电ICD电池。