首页> 外文会议>Conference on photonics applications in astronomy, communications, industry, and high-energy physics experiments >Design and construction of the Artificial Patient module for testing bioimpedance measuring devices
【24h】

Design and construction of the Artificial Patient module for testing bioimpedance measuring devices

机译:用于测试生物阻抗测量设备的人工患者模块的设计和构造

获取原文
获取外文期刊封面目录资料

摘要

The purpose of this study was to describe the design of the electronic module for testing bioimpedance measuring devices, for example impedance cardiographs or impedance pneumographs. Artificial Patient was conceived as an electronic equivalent of the impedance of skin-electrode interface and the impedance between electrodes - measured one. Different approaches in imitating a resistance of skin and an impedance of electrode-skin connection were presented. The module was adapted for frequently applied tetrapolar electrode configuration. Therefore the design do not enclose the elements simulating impedance between skin and receiver electrodes due to negligible effect of this impedance on the current flow through the receiver. The Artificial Patient enables testing either application generators, or receiver parts, particularly the level of noise and distortions of the signal. Use of digitally controlled potentiometer allows simulating different tissue resistances changes such as constant values, very-low-frequency and low-frequency changes corresponding to those caused by breathing or heart activity. Also it allows distorting signals in order to test algorithms of artifacts attenuation.
机译:这项研究的目的是描述用于测试生物阻抗测量设备的电子模块的设计,例如阻抗心动描记器或阻抗呼吸描记器。人工患者被认为是皮肤-电极界面的阻抗和电极之间的阻抗的电子等效物-测量一个。提出了模仿皮肤电阻和电极-皮肤连接阻抗的不同方法。该模块适用于经常使用的四极电极配置。因此,由于该阻抗对流经接收器的电流的影响可忽略不计,因此该设计并未在皮肤和接收器电极之间封装模拟阻抗的元件。人工患者可以测试应用程序生成器或接收器部件,尤其是噪声水平和信号失真。使用数字电位器可以模拟不同的组织电阻变化,例如恒定值,与呼吸或心脏活动引起的变化相对应的极低频和低频变化。它还允许失真信号,以便测试伪像衰减算法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com