Dose Reduction in Fluoroscopic Interventions Using a Combination of a Region of Interest (ROI) X-Ray Attenuator and Spatially-Different, Temporally-Variable Temporal Filtering

机译:结合使用感兴趣区域(ROI)X射线衰减器和时空时空变化的时空滤波技术,在荧光镜干预中减少剂量

获取原文

摘要

A novel dose reduction technique for fluoroscopic interventions involving a combination of a material x-ray region of interest (ROI) attenuator and spatially different, temporally variable ROI temporal recursive filter, was used to guide the catheter to the ROI in three live animal studies, two involving rabbits and one involving a sheep. In the two rabbit studies presented , a catheter was guided to the entrance of the carotid artery. With the added ROI attenuator the image under the high attenuation region is very noisy. By using temporal filtering with a filter weight of 0.6 on previous frames, the noise is reduced. In the sheep study the catheter was guided to the descending aorta of the animal. The sheep offered a relatively higher attenuation to the incident x-rays and thus a higher temporal filter weight of 0.8 on previous frames was used during the procedure to reduce the noise to levels acceptable by the interventionalist. The image sequences from both studies show that significant dose reduction of 5-6 times can be achieved with acceptable image quality outside the ROI by using the above mentioned technique. Even though the temporal filter weighting outside the ROI is higher, the consequent lag does not prevent perception of catheter movement.
机译:用于荧光透视干预的新剂量还原技术,涉及利益物质X射线区域(ROI)衰减器和空间不同,时间可变ROI时间递归过滤器的组合,用于将导管引导到三个活的动物研究中的ROI,两个涉及兔子和一个涉及羊的人。在提出的两种兔子研究中,导管被引导到颈动脉的入口。随着添加的ROI衰减器,高衰减区域下的图像非常嘈杂。通过在先前帧上使用滤波器重量为0.6的时间滤波,减少了噪声。在绵羊研究中,导管被引导到动物的下降主动脉。该绵羊向入射X射线提供相对较高的衰减,因此在过程中使用更高的时间滤波器重量为0.8,以减少涉及介入者可接受的噪声。两种研究的图像序列表明,通过使用上述技术可以通过ROI外部的可接受的图像质量实现5-6倍的显着剂量降低。即使ROI外部的时间滤波器较高,随后的滞后也不会阻止导管运动的感知。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号