Comparison of the Radial Compliance of Three Kinds of Vascular Prostheses and Pig's Carotid Based on Dynamic-Simulated Condition

机译:基于动态模拟条件的三种血管假体和猪颈瘤径向依从性的比较

获取原文
获取外文期刊封面目录资料

摘要

There were some problems in the vascular prostheses, which effected the health of the patients. This paper aims to presenting the radial compliance of three kinds of vascular prostheses and the pig's carotid based on the dynamic-simulated condition. The radial compliance of knitted structure vascular prostheses with crimps ranged from 2.52 to 0.94%/100mmHg. It was superior to e-PTFE and woven prostheses', which ranged from 0.73 to 0.31%/100mmHg and from 0.81 to 0.22%/100mmHg. But comparing with the radial compliance of pig's carotid, which ranged from 7.21 to 10.04%/100mmHg, there was a big gap between them. And also the trend of the radial compliance of vascular prostheses was different with the real pig's carotid. There is a lot of work left to improve the radial compliance. Not only to change the vascular prostheses' compliance, but also to meet with the real vascular.
机译:血管假体存在一些问题,这影响了患者的健康。本文旨在基于动态模拟条件呈现三种血管假体和猪颈动脉的径向依从性。针织结构血管假体的径向顺应性为卷曲的血管假体为2.52至0.94%/ 100mmHg。它优于E-PTFE和编织假体,其范围为0.73至0.31%/ 100mmHg,0.81至0.22%/ 100mmHg。但与猪颈动脉的径向顺应相比,从7.21到10.04%/ 100mmhg,它们之间存在很大差距。而且血管假肢的径向顺应性的趋势与真正的猪的颈动脉不同。还有很多工作来提高径向合规性。不仅要改变血管假肢的合规性,还要与真正的血管相遇。